TOC made easy -大昌华嘉商业(中国)有限公司

大昌华嘉咨询热线:400-821-0778

您现在的位置:首页 >服务 >TOC made easy

TOC made easy 发布时间:2014/05/27 14:13:55

TOC made easy Concept The concept of the high temperature TOC analyzer liquiTOC includes a separation of matrix compounds like salts or non combustible particles which is important to protect the catalyst from being poisoned or blocked. On the other hand a dynamic heating process allows high sample volumes up to 4ml thus giving a high sensitivity for low level TOC measurements. The basic concept of the instrument is shown in fig. 1. Summary The concept of matrix separation offers new applications for routine TOC analysis. The dynamic sample heating enables not only an effective matrix separation, it is as well possible to run high sample volumes up to 4 ml which was beyond the possibilities for high temperature TOC analyzer up to now. References [1] Verordnung über die Qualität von Wasser für den menschlichen Gebrauch –TrinkwV 2001, Bundesministerium für Gesundheit [2] Guidelines for the determination of total organic carbon (TOC) and dissolved organic carbon (DOC) in water, EN 1484, European Committee for Standardization, 1997

点击索取

您索取的服务为: